ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO
Report
- Report Number
- 2953200-2011-00921
- Event Type
- Injury
- Date Received
- April 29, 2011
- Date of Event
- April 1, 2011
- Report Date
- April 1, 2011
- Manufacturer
- MEDTRONIC CARDIOVASCULAR
- Product Code
- MIH
- PMA / PMN Number
- P990020
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVAL, RESULTS: (STENT GRAFT MIGRATION, RUPTURE). RESULTS/CONCLUSION: (SHORT AND FUNNEL-SHAPED AORTIC NECK).
AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROX 26 MONTHS AGO. ANEURYSM AND VESSEL MORPHOLOGY AT THE TIME OF IMPLANT ARE UNK. IT WAS REPORTED THAT AT THE ONE YEAR FOLLOW-UP A DISTAL MIGRATION OF APPROX ONE CM WAS NOTED BUT LEFT INCORRECT. THE PT PRESENTED EMERGENTLY APPROX ONE MONTH AGO WITH LEG, BACK, AND ABDOMINAL PAIN AND THE BIFURCATED STENT GRAFT WAS FOUND TO BE 20 MM BELOW THE RENAL ARTERIES. THE SYMPTOMS AND THE CT SCAN SEEMED TO SUGGEST A RUPTURED ANEURYSM; HOWEVER, THIS COULD NOT BE CONFIRMED AS THERE WAS NO EXTRAVASATION FROM THE ANEURYSM SAC. THERE WAS NO TYPE I OR OTHER ENDOLEAK EVIDENT. AT THE TIME OF MIGRATION, THE ANEURYSM WAS 5-6 CM IN DIAMETER. THE AORTIC NECK WAS REVERSE FUNNEL-SHAPED VARYING IN DIAMETER FROM 23-24-28 MM OVER A LENGTH OF 10 MM OR LESS WITHOUT ANGULATION. THE NECK WAS QUITE SHORT AT 10 MM OR LESS, WHICH IS THOUGHT TO HAVE CONTRIBUTED TO THE MIGRATION. A 28X28X49 MM ENDURANT PROXIMAL CUFF WAS PLACED TO RESOLVE THE MIGRATION SUCCESSFULLY. NO ADD'L CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO | MIH | MEDTRONIC CARDIOVASCULAR | NA | V00276107 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Required Intervention |