FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO

MDR report key: 2091664 · Received April 29, 2011

Report

Report Number
2953200-2011-00912
Event Type
Injury
Date Received
April 29, 2011
Date of Event
March 31, 2011
Report Date
March 31, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR SANTA ROSA
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL, RESULTS: (STENT GRAFT MIGRATION, ENDOLEAK). RESULTS/CONCLUSIONS: (DILATED AORTIC NECK).

Description of Event or Problem · 1

A TALENT STENT GRAFT WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROX TWO YRS AGO. THE TALENT BIFURCATED STENT GRAFT WAS ORIGINALLY DEPLOYED ONE CM BELOW THE LEVEL OF THE RENAL ARTERIES. THE NATIVE AORTIC NECK WAS 26.5 MM IN DIAMETER AND THERE WAS A PROXIMAL TYPE I ENDOLEAK AND NO INTERVENTION WAS DONE AT THE TIME. IT WAS REPORTED THAT THE PT PRESENTED EMERGENTLY APPROX ONE MONTH AGO. A CT SCAN WAS PERFORMED AND DEMONSTRATED THAT THE TALENT BIFURCATED STENT GRAFT HAS MIGRATED 4 CM FROM THE RENAL ARTERY WITH A TYPE I ENDOLEAK PRESENT. CURRENTLY AT THE RENAL ARTERIES THE NATIVE AORTA MEASURED 33 MM IN DIAMETER, 34 MM IN DIAMETER AND 36 MM IN DIAMETER, USING 5MM SLICES. A 36 MM DIAMETER TALENT AORTIC CUFF WAS IMPLANTED; HOWEVER, IT WAS NOT LONG ENOUGH; THEREFORE, THE DECISION WAS MADE TO IMPLANT A TALENT THORACIC CAPTIVIA (REF MFR # 2953200-2011-00913). PRIOR TO INSERTION OF THE TALENT THORACIC CAPTIVIA THE STENT GRAFT WAS DEPLOYED ON THE BACK TABLE. THE PHYSICIAN CUT THE STENT GRAFT TO THE DESIRED LENGTH AND CREATED A FENESTRATION IN THE STENT GRAFT THEN RE-LOADED IT INTO THE DELIVERY CATHETER AND SUCCESSFULLY IMPLANTED IT IN THE PT. NO ADD'L CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC CARDIOVASCULAR SANTA ROSA NA V00195939

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention