FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2091572 · Received May 17, 2011

Report

Report Number
6000001-2011-04125
Event Type
Malfunction
Date Received
May 17, 2011
Date of Event
April 1, 2011
Report Date
April 28, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K982102
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE CONTAINING APPROXIMATELY 50 ML OF SOLUTION IN THE RESERVOIR. THE REPORTED CONDITION OF NO FLOW/NON-DELIVERY WAS NOT CONFIRMED. VISUAL EXAMINATION OF THE DEVICE SHOWED NO SIGNS OF BLOCKAGE THAT COULD HAVE CAUSED THE REPORTED CONDITION. TO VERIFY THE REPORTED CONDITION, THE RED LUER CAP WAS REMOVED. FLOW WAS READILY OBSERVED AT THE LUER AND CONTINUED TO FLOW WITHOUT STOPPING UNTIL THE SOLUTION WAS EMPTIED FROM THE RESERVOIR. NO ROOT CAUSE WAS IDENTIFIED, AS NO EVIDENCE OF PRODUCT MALFUNCTION WAS OBSERVED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. ADDITIONAL INFORMATION: PER REVIEW OF THE BATCH RECORDS, NO NONCONFORMANCE REPORT WAS DOCUMENTED FOR THIS LOT. ALL RELEASE CRITERIA WERE MET FOR THE BUILD OF THE LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE HAS BEEN RECEIVED AND IS AVAILABLE FOR EVALUATION. ONCE THE SAMPLE HAS BEEN EVALUATED A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A COMPLAINT THAT ONE HALF DAY INFUSOR, 5ML/HR 12PK DEVICE REPORTEDLY EXPERIENCED A NO-FLOW. THE CAP WAS REMOVED AND NO SOLUTION CAME OUT. THIS WAS PRIOR TO PATIENT CONNECTION. NO PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION. THIS WILL REFERENCE REPORT 1 OF 2 FROM THE FACILITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 10G003

Patients

Seq Age Sex Outcome Treatment
1