FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION¿ S8 SYSTEM

MDR report key: 20915696 · Received December 12, 2024

Report

Report Number
1723170-2024-03876
Event Type
Malfunction
Date Received
December 12, 2024
Date of Event
December 3, 2024
Report Date
January 3, 2025
Manufacturer
MEDTRONIC NAVIGATION, INC
Product Code
HAW
UDI-DI
00643169722187
PMA / PMN Number
K162309
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: PRODUCT ID 9735670 (SERIAL: (B)(6) ); PRODUCT TYPE: ; IMPLANT DATE N/A; EXPLANT DATE N/A SECTION D REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER MEDICAL PRODUCTS IN USE DURING THE EVENT INCLUDE: CAMERA 9735821 VEGA BASE S8 SVC H3) NO PARTS HAVE BEEN RETURNED FOR ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

WORK ORDER COMPLETED: G3, H6: A MANUFACTURER REPRESENTATIVE WENT TO THE SITE TO TEST THE NAVIGATION SYSTEM. IT WAS REPORTED THAT THERE WAS NO ISSUE. CODES B01, C19 AND D14 ARE APPLICABLE TO THIS EVALUATION. H3: ANALYSIS WAS PERFORMED FOR PRODUCT 9735821, LOT NUMBER P902695. IT WAS REPORTED THAT THE RETURNED PSU HAD MANY NICKS AND SCRATCHES ON THE HOUSING AND LENSES. A CHECK OF THE EVENT LOG SHOWED INTERMITTENT FIRMWARE INCOMPATIBILITY AND INTERMITTENT ILLUMINATOR CURRENT LOW. THERE WAS A BATTERY VOLTAGE LOW MESSAGE ALONG WITH BUMP DETECTED AND STORAGE TEMPERATURE EXCEEDED. THE PSU FAILED AN ACCURACY TEST (AAK) AT .484 MM WITH A PASSING THRESHOLD OF .250 MM. CODES B01, C08 AND D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THERE WAS A LOCALIZER FAULTED ERROR ON THE CAMERA CART. THEY SWAPPED THE CAMERA CART TO A DIFFERENT MAIN CART TO ENSURE THERE WAS NO ISSUE WITH THE CABLING TO THE MAIN CART. THE LOCALIZER FAULTED MESSAGE STILL APPEARED WHEN CONNECTED TO A NEW MAIN CART. THE REPRESENTATIVE STATED THE GREEN AND AMBER LIGHT WERE ILLUMINATED. THE REPRESENTATIVE ACCESSED THE NDITOOLBOX AND A BUMP AND INTERNAL TEMPERATURE STATUSES WERE TRIGGERED. THE STATUSES MESSAGES REPORT "BUMP DETECTOR BATTERY FAULT. RETURN FOR SERVICE." THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2204145 STEALTHSTATION¿ S8 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC 9735665 00643169722187

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown SEE H11...