FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2091547 · Received May 17, 2011

Report

Report Number
2531779-2011-03461
Event Type
Malfunction
Date Received
May 17, 2011
Date of Event
April 24, 2011
Report Date
April 24, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: NO ALARMS RELATED TO THE COMPLAINT WERE OBSERVED IN THE ALARM HISTORY. A CRACK WAS OBSERVED IN THE BATTERY COMPARTMENT. THE THREADS ON THE BATTERY CAP WERE OBSERVED TO BE STRIPPED. THE BATTERY CAP WAS UNABLE TO ESTABLISH AN ELECTRICAL CONNECTION, INABILITY TO POWER PUMP DUPLICATED. A TEST CAP WAS USED FOR TESTING PURPOSES. THE PUMP POWERS ON NORMAL WITH AUDIO, DISPLAY, AND BACKLIGHT WORKING. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER ISSUES OCCURRING. A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP WAS PERFORMED AND THE PUMP WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. SOFTWARE VERSION: B2J, DATE OF MANUFACTURE: 6/2008.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT SHE NOTICED THE PUMP WAS OFF AFTER TAKING A SHOWER. SHE NOTED THAT THE DISPLAY SCREEN WAS BLANK, AND THERE WERE NO TONES OR VIBRATIONS PRESENT. THE PATIENT DENIED EXPOSING THE PUMP TO WATER. SHE CONFIRMED THAT SHE WAS USING AN ENERGIZER ULTIMATE LITHIUM BATTERY. THE PATIENT STATED THAT SHE CHANGED THE BATTERY, BUT THE ISSUE DID NOT RESOLVE; SHE NOTED THAT THE PUMP DID NOT MAINTAIN TIME AND DATE SETTINGS WITH THE BATTERY CHANGE. SHE NOTED THERE WAS NO VISIBLE CORROSION IN THE BATTERY COMPARTMENT, AND THERE WAS NO STRUCTURAL DAMAGE TO THE BATTERY CAP OR TO THE BATTERY COMPARTMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 1250 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 60 YR