ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-03461
- Event Type
- Malfunction
- Date Received
- May 17, 2011
- Date of Event
- April 24, 2011
- Report Date
- April 24, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
(B)(4). DEVICE EVALUATION: THE PUMP WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: NO ALARMS RELATED TO THE COMPLAINT WERE OBSERVED IN THE ALARM HISTORY. A CRACK WAS OBSERVED IN THE BATTERY COMPARTMENT. THE THREADS ON THE BATTERY CAP WERE OBSERVED TO BE STRIPPED. THE BATTERY CAP WAS UNABLE TO ESTABLISH AN ELECTRICAL CONNECTION, INABILITY TO POWER PUMP DUPLICATED. A TEST CAP WAS USED FOR TESTING PURPOSES. THE PUMP POWERS ON NORMAL WITH AUDIO, DISPLAY, AND BACKLIGHT WORKING. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER ISSUES OCCURRING. A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP WAS PERFORMED AND THE PUMP WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. SOFTWARE VERSION: B2J, DATE OF MANUFACTURE: 6/2008.
THE PATIENT REPORTED THAT SHE NOTICED THE PUMP WAS OFF AFTER TAKING A SHOWER. SHE NOTED THAT THE DISPLAY SCREEN WAS BLANK, AND THERE WERE NO TONES OR VIBRATIONS PRESENT. THE PATIENT DENIED EXPOSING THE PUMP TO WATER. SHE CONFIRMED THAT SHE WAS USING AN ENERGIZER ULTIMATE LITHIUM BATTERY. THE PATIENT STATED THAT SHE CHANGED THE BATTERY, BUT THE ISSUE DID NOT RESOLVE; SHE NOTED THAT THE PUMP DID NOT MAINTAIN TIME AND DATE SETTINGS WITH THE BATTERY CHANGE. SHE NOTED THERE WAS NO VISIBLE CORROSION IN THE BATTERY COMPARTMENT, AND THERE WAS NO STRUCTURAL DAMAGE TO THE BATTERY CAP OR TO THE BATTERY COMPARTMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 1250 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |