FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2091536 · Received May 17, 2011

Report

Report Number
2531779-2011-03460
Event Type
Malfunction
Date Received
May 17, 2011
Date of Event
April 19, 2011
Report Date
April 19, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, IT WAS FOUND THAT REBOOTS WERE OBSERVED IN THE PUMP DOWNLOAD HISTORY. THERE WERE NO ALARMS RELATED TO THE COMPLAINT OBSERVED IN THE ALARM HISTORY. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO ALARMS OR POWER ISSUES OCCURRING. THE PUMP WAS OPENED AND CORROSION WAS FOUND ON THE PCB. A CRACKED BATTERY COMPARTMENT WAS FOUND AS WELL AS BATTERY COMPARTMENT LEAK AND MOISTURE IN THE PUMP. THERE WAS NO DEFECT FOUND.

Additional Manufacturer Narrative · 1

AFTER THE EVALUATION IS COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILLED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THIS COMPLAINT IS BEING REPORTED AS A MALFUNCTION DUE TO THE ALLEGED SELF-REBOOT ISSUE. THERE WAS NO EVIDENCE OF PRODUCT MISUSE OR TRAUMA TO THE DEVICE. THE REPORTER DENIES ANY BLOOD GLUCOSE EXCURSION AS A RESULT OF THE PRODUCT ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 51 YR