FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2091520 · Received May 17, 2011

Report

Report Number
6000001-2011-04119
Event Type
Malfunction
Date Received
May 17, 2011
Date of Event
March 31, 2011
Report Date
April 26, 2011
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K883577
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE CONDITION OF AN INFUSO.R. WITH A "DELAY"ISSUE WAS CONFIRMED AND REPRODUCED DURING PRODUCT EVALUATION. THE ASSIGNABLE CAUSE WAS DETERMINED TO BE A DEFECTIVE REED SWITCH. THE REED SWITCH WAS REPLACED IN ORDER TO FIX THE REPORTED CONDITION. ADDITIONAL INFORMATION: A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THIS PUMP HAS NEVER PREVIOUSLY BEEN SENT TO BAXTER FOR SERVICE AND, THEREFORE, THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS REPORTED PROBLEM FOR THIS PUMP.

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER TO REPORT AN INFUSOR PUMP IN WHICH THERE IS A DELAY. THERE IS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. IT IS UNKNOWN WHEN THE EVENT OCCURRED. THE EVENT OCCURRED IN THE OPERATING ROOM. THERE IS NOT AN ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSOR PUMP, INFUSION FRN BAXTER HEALTHCARE CORPORATION

Patients

Seq Age Sex Outcome Treatment
1