FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2091415 · Received May 6, 2011

Report

Report Number
3002158293-2011-00522
Event Type
Malfunction
Date Received
May 6, 2011
Date of Event
April 2, 2011
Report Date
May 5, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 107) HAS BEEN CONFIRMED. UPON EVAL THE ELECTRODE BELT CONNECTOR WAS BROKEN, EXPOSING WIRES. THE CAUSE FOR THE BROKEN TRUNK CONNECTOR CANNOT BE POSITIVELY DETERMINED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE BROKEN CONNECTOR. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS MONITOR IS DISPLAYING SERVICE CODE 107. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT AND MONITOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 55 YR