FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES
MDR report key: 2091306
·
Received May 4, 2011
Report
- Report Number
- 3015876-2011-00365
- Event Type
- Malfunction
- Date Received
- May 4, 2011
- Date of Event
- April 4, 2011
- Report Date
- April 4, 2011
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- MKJ
- PMA / PMN Number
- K991910
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED ALL FOUR BATTERY CONNECTOR PINS AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WOULD POWER ON/OFF BY ITSELF. THERE WAS NO PT USE ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES | MKJ | PHYSIO-CONTROL, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |