FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2091302 · Received May 17, 2011

Report

Report Number
2531779-2011-03451
Event Type
Malfunction
Date Received
May 17, 2011
Report Date
April 23, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE TOTAL DAILY DOSE HISTORY WAS REVIEWED FROM (B)(6) 2010 TO THE END OF PUMP USE ON (B)(6) 2011 AND THE DAILY INSULIN DELIVERY TOTALS CORRECTLY REFLECT THE USERS PROGRAMMED BASAL RATES. THE PUMP HISTORY SHOWS A 'REPLACE BATTERY' ALARM ON (B)(6) 2011 18:13. DURING INVESTIGATION, UNABLE TO REPRODUCE SECONDARY ERROR. THE PUMP ACCURATELY DELIVERS 2.00 UNITS/HR FOR 29-HR PERIOD. DHR REVIEW: DATE OF MANUFACTURE: 2009-11, SOFTWARE VERSION/REVISION: E3A, WITHIN SPECIFICATIONS WHEN RELEASED: YES, DISCREPANCIES IDENTIFIED: NO, COMMENTS: NO.

Description of Event or Problem · 1

ON (B)(6) 2011 IT WAS REPORTED THE PATIENT OBTAINED BATTERY ALARMS AND THEN NO POWER ON THE REPORTED PUMP. THE PATIENT REPLACED THE BATTERIES TO NO AVAIL; THE PUMP GAVE ALARMS AND THEN POWERED OFF. THE PATIENT DID NOT EXPERIENCE ANY SYMPTOMS OF HIGH OR LOW BLOOD GLUCOSE LEVELS AND DID NOT SEEK ANY MEDICAL ATTENTION DUE TO THIS ISSUE. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION OTP GLUCOSE MGMT SYSTEM

Patients

Seq Age Sex Outcome Treatment
1 51 YR