FDA Adverse Event Malfunction Summary report: N

VS3: NBP, SP02, TEMP

MDR report key: 2091294 · Received April 20, 2011

Report

Report Number
1218950-2011-01049
Event Type
Malfunction
Date Received
April 20, 2011
Report Date
March 24, 2011
Manufacturer
PHILIPS HEALTHCARE
Product Code
MHX
PMA / PMN Number
K082280
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTED DESCRIPTION NOTES THAT A TECHNICAL MESSAGE WAS DISPLAYED ON THE PATIENT MONITOR "SPEAKER MALFUNCTION" WITH NO AUDIO FUNCTION. ACCORDING TO THE CUSTOMER, THE SPEAKER WAS REPLACED INITIALLY AFTER FINDING THAT NO AUDIO WAS AVAILABLE FROM THE MONITOR. THIS REPLACEMENT DID NOT RESOLVE THE AUDIO ISSUE. THE CUSTOMER REPLACED THE MONITOR'S MAINBOARD, AND THE SPEAKER MALFUNCTION WAS RESOLVED. THE MONITOR WAS TESTED AND RETURNED TO USE WITHIN THE HOSPITAL. NO FURTHER INVESTIGATION OR ACTION IS WARRANTED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A TECHNICAL MESSAGE WAS DISPLAYED ON THE PATIENT MONITOR "SPEAKER MALFUNCTION" AND THERE WAS NO AUDIO FUNCTION. NO PATIENT HARM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VS3: NBP, SP02, TEMP MHX PHILIPS HEALTHCARE 863073

Patients

Seq Age Sex Outcome Treatment
1