FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2091263 · Received April 20, 2011

Report

Report Number
2531779-2011-02731
Event Type
Malfunction
Date Received
April 20, 2011
Report Date
March 27, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE DOWN ARROW AND OK BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER ALL BUTTON CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BUTTONS ARE REQUIRING SEVERAL PRESSES. IT WAS REPORTED THERE ARE NO HOLES OR TEARS IN THE KEYPAD, BUT STATED THAT THE KEYPAD IS VERY WORN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1