FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 2091262
·
Received April 20, 2011
Report
- Report Number
- 2531779-2011-02730
- Event Type
- Malfunction
- Date Received
- April 20, 2011
- Report Date
- March 27, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS PEELED AT THE OK BUTTON. ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. THE DOWN ARROW AND OK KEY CONTACTS WERE INVERTED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD IS UNRESPONSIVE. THE PATIENT STATED, IT STARTED (B)(6) AGO. SHE REPORTED THAT TONIGHT IS WHEN SHE NOTICED THAT THE KEYPAD IS PEELING AND THAT THERE IS A TEAR AT THE UP AND DOWN ARROWS DUE TO PRESSING HARD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |