FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2091262 · Received April 20, 2011

Report

Report Number
2531779-2011-02730
Event Type
Malfunction
Date Received
April 20, 2011
Report Date
March 27, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS PEELED AT THE OK BUTTON. ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. THE DOWN ARROW AND OK KEY CONTACTS WERE INVERTED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD IS UNRESPONSIVE. THE PATIENT STATED, IT STARTED (B)(6) AGO. SHE REPORTED THAT TONIGHT IS WHEN SHE NOTICED THAT THE KEYPAD IS PEELING AND THAT THERE IS A TEAR AT THE UP AND DOWN ARROWS DUE TO PRESSING HARD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA