FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2091252 · Received April 20, 2011

Report

Report Number
2531779-2011-02722
Event Type
Malfunction
Date Received
April 20, 2011
Date of Event
March 20, 2011
Report Date
March 24, 2011
Manufacturer
ANIMAS CORP
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. ALL KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE DURING TESTING. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE STICKING AND INTERMITTENTLY UNRESPONSIVE. THE PT REPORTED THAT THE DOWN ARROW AND OK KEYPAD BUTTONS HAVE BEEN INTERMITTENTLY UNRESPONSIVE AND STICKING FOR THE PAST 4 TO 5 DAYS. SHE CONFIRMED THAT THE KEYPAD IS INTACT; DENIED EXPOSING THE PUMP TO MOISTURE; AND STATED THAT SHE WEARS THE PUMP IN HER BRA OR ON HER WAIST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 48 YR