ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 2091252
·
Received April 20, 2011
Report
- Report Number
- 2531779-2011-02722
- Event Type
- Malfunction
- Date Received
- April 20, 2011
- Date of Event
- March 20, 2011
- Report Date
- March 24, 2011
- Manufacturer
- ANIMAS CORP
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. ALL KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE DURING TESTING. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE STICKING AND INTERMITTENTLY UNRESPONSIVE. THE PT REPORTED THAT THE DOWN ARROW AND OK KEYPAD BUTTONS HAVE BEEN INTERMITTENTLY UNRESPONSIVE AND STICKING FOR THE PAST 4 TO 5 DAYS. SHE CONFIRMED THAT THE KEYPAD IS INTACT; DENIED EXPOSING THE PUMP TO MOISTURE; AND STATED THAT SHE WEARS THE PUMP IN HER BRA OR ON HER WAIST.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP | ONE TOUCH PING |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR |