FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2091248 · Received April 20, 2011

Report

Report Number
2531779-2011-02728
Event Type
Malfunction
Date Received
April 20, 2011
Report Date
March 27, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. DURING INVESTIGATION, ALL KEY CONTACTS WERE OBSERVED TO BE MISALIGNED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BUTTONS ARE STICKING. IT WAS REPORTED THAT THE PUMP IS NOT EXPOSED TO WATER AND THE KEYPAD IS INTACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1