ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
Report
- Report Number
- 2531779-2011-02723
- Event Type
- Malfunction
- Date Received
- April 20, 2011
- Date of Event
- March 3, 2011
- Report Date
- March 24, 2011
- Manufacturer
- ANIMAS CORP
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. ALL KEYPADS BUTTONS WERE INTERMITTENTLY RESPONSIVE DURING TESTING. IT WAS OBSERVED THAT THE KEYPAD BUTTONS DO NOT CLICK AND SPRING BACK WHEN PRESSED. DURING INVESTIGATION, THE UP ARROW AND OK KEY CONTACTS WERE OBSERVED TO BE MISALIGNED. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS.
IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE PT REPORTED THAT THE DOWN ARROW KEYPAD BUTTONS HAS BEEN INTERMITTENTLY UNRESPONSIVE FOR THE (B)(6) AND REQUIRES NUMEROUS PRESSES TO OBTAIN A RESPONSE. SHE NOTED THAT ALL KEYPAD BUTTONS CLICK AND SPRING BACK AS EXPECTED WHEN PRESSED. THE PT DENIED PHYSICAL DAMAGE TO THE RUBBER KEYPAD; DENIED EXPOSING THE PUMP TO WATER; AND STATED THAT SHE WEARS THE PUMP IN HER BRA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP | ONE TOUCH PING |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR |