FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2091238 · Received April 20, 2011

Report

Report Number
2531779-2011-02712
Event Type
Malfunction
Date Received
April 20, 2011
Date of Event
March 23, 2011
Report Date
March 23, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE CONTRAST, UP ARROW, AND OK KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE, AND THE DOWN ARROW BUTTONS WAS UNRESPONSIVE DURING TESTING. IT WAS OBSERVED THAT ALL BUTTONS SPRING BACK WHEN PRESSED EXCEPT THE DOWN ARROW BUTTON. DURING INVESTIGATION, THE DOWN ARROW KEY CONTACT WAS OBSERVED TO BE INVERTED. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE STICKING AND UNRESPONSIVE. A FAMILY MEMBER REPORTED THAT THE UP ARROW AND DOWN ARROW KEYPAD BUTTONS ARE STICKING AND ARE NOT RESPONDING TO PRESSES. SHE NOTED THAT THE BUTTONS DO NOT CLICK WHEN PRESSED. THE FAMILY MEMBER CONFIRMED THAT THE KEYPAD IS INTACT; AND DENIED EXPOSING THE PUMP TO WATER AND CLEANING PRODUCTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 74 YR