OLYMPUS AUXILIARY WATER TUBE
Report
- Report Number
- 8010047-2011-00090
- Event Type
- Other
- Date Received
- May 5, 2011
- Date of Event
- January 1, 2008
- Report Date
- April 6, 2011
- Manufacturer
- OLYMPUS MEDICAL SYSTEM CORPORATION
- Product Code
- FDS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NURSE
Narratives
OLYMPUS FOLLOWED UP WITH THE USER FACILITY TO OBTAIN ADD'L INFO REGARDING THE REPORT, AND WAS INFORMED THAT THE USER FACILITY WAS REPORTEDLY CLEANING THE MAJ-855 WITH A DETERGENT SOLUTION AS PER THE INSTRUCTIONS FOR USE, BUT THEN PLACING THE MAJ-855 AUXILIARY WATER CHANNEL IN THE AER WITHOUT CONNECTING THE TUBING TO THE FLUSHING PORT. THE AER MFR ALSO CONTRAINDICATES REPROCESSING OF THE MAJ-855 IN THE MODEL OF AER USED AT THE FACILITY. OLYMPUS WAS INFORMED THE USER FACILITY HAS SINCE IMPLEMENTED APPROPRIATE REPROCESSING METHODS. AS PART OF THE INVESTIGATION INTO THIS REPORT AN OLYMPUS' ENDOSCOPY SUPPORT SPECIALIST (ESS) HAVE BEEN DISPATCHED TO THE USER FACILITY TO F/U AND PROVIDE ANY NECESSARY IN-SERVICE TRAINING ON APPROPRIATE REPROCESSING METHODS. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.
OLYMPUS WAS INFORMED THAT THE USER FACILITY WAS NOT REPROCESSING THE AUXILIARY WATER TUBE IN ACCORDANCE WITH THE DIRECTIONS FOR USE. THE USER FACILITY PERSONNEL WERE REPORTEDLY PLACING THE AUXILIARY WATER TUBE INSIDE THE BASIN OF THE AUTOMATIC ENDOSCOPE REPROCESSOR (AER) WITHOUT CONNECTING THE MAJ-855 TO ANY OF THE FLUSHING CONNECTORS OF THE AER. IN THIS CONFIGURATION, IT IS UNLIKELY THAT THE INTERIOR OF THE MAJ-855 TUBING WAS BEING ADEQUATELY FLUSHED WITH DISINFECTANT. THIS PRACTICE WAS SAID TO HAVE BEEN ONGOING SINCE SOME TIME IN 2008. THERE WERE NO REPORTS OF ANY ADVERSE IMPACT TO PT OR USERS ASSOCIATED WITH THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPUS AUXILIARY WATER TUBE | AUXILIARY WATER TUBE | FDS | OLYMPUS MEDICAL SYSTEM CORPORATION | MAJ-855 | NA |