FDA Adverse Event Malfunction Summary report: N

CENTURY BED

MDR report key: 2091094 · Received May 12, 2011

Report

Report Number
1824206-2011-02672
Event Type
Malfunction
Date Received
May 12, 2011
Date of Event
April 19, 2011
Report Date
April 19, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BRAKE CABLE JAMMED UNDER THE BENT STIFFENER PLATE AND THE TOP OF THE BRAKE BLOCK WAS BROKEN. HE REPLACED THE STIFFENER PLATE, BEARINGS, TOP AND BOTTOM OF THE BRAKE BLOCK AND ADJUSTED THE BRAKE/STEER TO REPAIR THE BED.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THE BRAKES ARE NOT HOLDING ON THE BED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTURY BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 835

Patients

Seq Age Sex Outcome Treatment
1 UNK