Report Number
1824206-2011-02662
Event Type
Malfunction
Date Received
May 12, 2011
Date of Event
April 19, 2011
Report Date
April 19, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN REPLACED THE MANUAL TREND VALVE TO REPAIR THE BED.

Description of Event or Problem · 1

INFO REC'D INDICATES THE HEAD HI/LOW FUNCTION IS DRIFTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900