FDA Adverse Event
Malfunction
Summary report: N
POWER-PRO TL COT
MDR report key: 2091038
·
Received May 11, 2011
Report
- Report Number
- 1831750-2011-04513
- Event Type
- Malfunction
- Date Received
- May 11, 2011
- Date of Event
- April 12, 2011
- Report Date
- April 12, 2011
- Manufacturer
- STRYKER MEDICAL
- Product Code
- INK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
BOLT HOLES IN SUB FRAME. IT WAS DETERMINED THERE IS A POTENTIAL RISK TO SAFETY IF THE COT COMES OUT OF THE FASTENER SYSTEM WHEN LOADING BECAUSE THIS COULD POTENTIALLY RESULT IN INJURY TO THE PT AND/OR CAREGIVER. IT WAS REPORTED THAT ONE OF THE COTS WORKED ITSELF LOOSE FROM A FASTENER OVER REPEATED USE, HOWEVER THE CUSTOMER HAS 11 COTS AND IT CANNOT BE DETERMINED WHICH COT WAS INVOLVED IN THE REPORTED INCIDENT OF COMING LOOSE FROM THE FASTENER.
Description of Event or Problem · 1
IT WAS REPORTED THE COT CAME FREE FROM THE FASTENING SYSTEM. THERE WAS PT INVOLVEMENT, HOWEVER NO ADVERSE CONSEQUENCES HAVE BEEN REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | POWER-PRO TL COT | STRETCHER, WHEELED | INK | STRYKER MEDICAL | 6550 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |