FDA Adverse Event Injury Summary report: N

DEXTRUS 4136

MDR report key: 2090971 · Received May 3, 2011

Report

Report Number
1028232-2011-01005
Event Type
Injury
Date Received
May 3, 2011
Date of Event
March 15, 2011
Report Date
April 27, 2011
Manufacturer
BIOTRONIK SE & CO. KG,
Product Code
NVZ
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC REC'D INFORMATION THAT ONE DAY POST IMPLANT THE RIGHT VENTRICULAR (RV) LEAD WAS SUCCESSFULLY REPOSITIONED DUE TO LOSS OF CAPTURE AND DISLODGEMENT. THE LEAD REMAINS IMPLANTED IN THE PT. NO SPECIFIC MEASUREMENTS WERE PROVIDED AND NO ADVERSE PT EFFECTS WERE REPORTED. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEXTRUS 4136 PACER LEAD NVZ BIOTRONIK SE & CO. KG, 358754

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization