FDA Adverse Event
Malfunction
Summary report: N
MAESTRO DRILL
MDR report key: 2090698
·
Received April 19, 2011
Report
- Report Number
- 1811755-2011-01310
- Event Type
- Malfunction
- Date Received
- April 19, 2011
- Date of Event
- February 24, 2011
- Report Date
- March 24, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- HBB
- PMA / PMN Number
- K041754
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DRILL WAS RECEIVED BY THE MFR, HOWEVER THE COMPLAINT COULD NOT BE REPLICATED. THE DEVICE EVAL REVEALED THAT THE DRILL PERFORMED ACCORDING TO ALL SPECS, AND NO DAMAGE TO THE HOSE WAS DETECTED. DUE TO THE AGE OF THE DEVICE, THE HOSE ASSEMBLY WAS REPLACED FOR PREVENTATIVE MAINTENANCE PURPOSES. THE DRILL WAS RETURNED THE USER FACILITY.
Description of Event or Problem · 1
IT WAS REPORTED BY THE USER FACILITY THAT THE PNEUMATIC DRILL WAS LEAKING AIR. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAESTRO DRILL | HBB | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |