FDA Adverse Event Malfunction Summary report: N

MAESTRO DRILL

MDR report key: 2090698 · Received April 19, 2011

Report

Report Number
1811755-2011-01310
Event Type
Malfunction
Date Received
April 19, 2011
Date of Event
February 24, 2011
Report Date
March 24, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DRILL WAS RECEIVED BY THE MFR, HOWEVER THE COMPLAINT COULD NOT BE REPLICATED. THE DEVICE EVAL REVEALED THAT THE DRILL PERFORMED ACCORDING TO ALL SPECS, AND NO DAMAGE TO THE HOSE WAS DETECTED. DUE TO THE AGE OF THE DEVICE, THE HOSE ASSEMBLY WAS REPLACED FOR PREVENTATIVE MAINTENANCE PURPOSES. THE DRILL WAS RETURNED THE USER FACILITY.

Description of Event or Problem · 1

IT WAS REPORTED BY THE USER FACILITY THAT THE PNEUMATIC DRILL WAS LEAKING AIR. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAESTRO DRILL HBB STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK