FDA Adverse Event
Malfunction
Summary report: N
MICRO SAGITTAL SAW
MDR report key: 2090671
·
Received April 19, 2011
Report
- Report Number
- 1811755-2011-01298
- Event Type
- Malfunction
- Date Received
- April 19, 2011
- Date of Event
- March 21, 2011
- Report Date
- March 21, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
DURING QUALITY INVESTIGATION, THE CUSTOMER'S COMPLAINT WAS DUPLICATED. UPON DISASSEMBLY OF THE DEVICE, A DAMAGED PRESS PLUG, MOTOR, ROTOR, BEARINGS AND OTHER COMPONENT PARTS WERE NOTED. THE FAULTY PARTS WERE REPLACED AND THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
A STRYKER SAGITTAL SAW WAS SENT IN FOR REPAIR DUE TO OVERHEATING. THERE WAS NO PT INVOLVEMENT; NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICRO SAGITTAL SAW | ERL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |