Additional Manufacturer Narrative · 1

DURING QUALITY INVESTIGATION, THE CUSTOMER'S COMPLAINT WAS DUPLICATED. UPON DISASSEMBLY OF THE DEVICE, A DAMAGED PRESS PLUG, MOTOR, ROTOR, BEARINGS AND OTHER COMPONENT PARTS WERE NOTED. THE FAULTY PARTS WERE REPLACED AND THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

A STRYKER SAGITTAL SAW WAS SENT IN FOR REPAIR DUE TO OVERHEATING. THERE WAS NO PT INVOLVEMENT; NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.