FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 2090670
·
Received April 19, 2011
Report
- Report Number
- 1811755-2011-01297
- Event Type
- Malfunction
- Date Received
- April 19, 2011
- Date of Event
- March 16, 2011
- Report Date
- March 23, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DURING QUALITY INVESTIGATION, THE CUSTOMER'S COMPLAINT WAS CONFIRMED; THE DEVICE EXCEEDED THE MAXIMUM ALLOWED TEMPERATURE RISE DURING TESTING. UPON DISASSEMBLY OF THE DEVICE, CORROSION DAMAGE WAS NOTED ON THE MOTOR, ROTOR AND DRIVESHAFT. THE PROBABLE CAUSE OF THE FAILURE WAS ATTRIBUTED TO DAMAGED SPINDLE HOUSING, BURNT MOTOR CARTRIDGE PINS, WORN ROTOR AND DRIVESHAFT BEARINGS. THE DAMAGED PARTS WERE REPLACED AND THE DEVICE REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
A STRYKER CORE IMPACTION DRILL WAS SENT IN FOR REPAIR DUE TO OVERHEATING DURING A QUALITY CHECK. THERE WAS NO PT INVOLVEMENT; NO ADVERSE CONSEQUENCE WAS ASSOCIATED WITH THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |