FDA Adverse Event
Malfunction
Summary report: N
CORE MICRO DRILL
MDR report key: 2090666
·
Received April 19, 2011
Report
- Report Number
- 1811755-2011-01303
- Event Type
- Malfunction
- Date Received
- April 19, 2011
- Date of Event
- March 18, 2011
- Report Date
- March 21, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- HBE
- PMA / PMN Number
- K032303
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
DURING FAILURE ANALYSIS, THE CUSTOMER'S COMPLAINT WAS DUPLICATED; THE DEVICE OVERHEATED DURING TESTING. FURTHER INVESTIGATION REVEALED CORROSION DAMAGE WITHIN THE MOTOR, ROTOR, PRESS PLUG AND OTHER COMPONENT PARTS. ACCORDING TO THE ANALYSIS NOTES, THE ROTOR WAS STUCK IN THE MOTOR DUE TO CORROSION; THIS INTERFERENCE MAY CAUSE FRICTION WHICH COULD LEAD TO THE OVERHEATING DESCRIBED BY THE CUSTOMER. THE DAMAGED PARTS WERE REPLACED AND THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
A STRYKER CORE MICRO DRILL WAS SENT IN FOR REPAIR DUE TO OVERHEATING DURING A PROCEDURE. NO ADVERSE CONSEQUENCE WAS REPORTED; ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT ANY PROCEDURE DELAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE MICRO DRILL | HBE | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |