FDA Adverse Event Malfunction Summary report: N

MICRO SAGITTAL SAW

MDR report key: 2090665 · Received April 19, 2011

Report

Report Number
1811755-2011-01302
Event Type
Malfunction
Date Received
April 19, 2011
Date of Event
March 22, 2011
Report Date
March 22, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

QUALITY INVESTIGATION REVEALED CORROSION DAMAGE THROUGHOUT THE INTERNAL PARTS OF THE DEVICE. THE PRESS PLUG, MOTOR, ROTOR, BEARINGS AND OTHER COMPONENT PARTS WERE DAMAGED. THE DAMAGED PARTS WERE REPLACED AND THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

A STRYKER SAGITTAL SAW WAS SENT IN FOR REPAIR DUE TO OVERHEATING DURING A PROCEDURE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER DEVICE; NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO SAGITTAL SAW HBE STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK