FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2090605
·
Received April 26, 2011
Report
- Report Number
- 1720753-2011-06580
- Event Type
- Malfunction
- Date Received
- April 26, 2011
- Date of Event
- March 30, 2011
- Report Date
- April 26, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE GE REP CONDUCTED AN ONSITE INVESTIGATION. THE GENERAL PURPOSE OPERATING SYSTEM PCB WAS REPLACED AND THE SOFTWARE WAS RELOADED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE 9900 SYSTEM DISPLAYED A FILE SYSTEM CORRUPTION DETECTED ERROR MESSAGE. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |