FDA Adverse Event
Malfunction
Summary report: N
6800
MDR report key: 2090576
·
Received April 26, 2011
Report
- Report Number
- 1720753-2011-06532
- Event Type
- Malfunction
- Date Received
- April 26, 2011
- Date of Event
- April 1, 2011
- Report Date
- April 26, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE MANUFACTURER'S SERVICE REPRESENTATIVE PERFORMED AN ON-SITE INVESTIGATION. THE C-MOS SETTINGS WERE RESET AND THE BATTERY WAS REPLACED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE LEFT MONITOR OF THE 6800 SYSTEM WOULD NOT DISPLAY AN IMAGE. NO PATIENT INJURY WAS REPORTED.