FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 2090495 · Received April 26, 2011

Report

Report Number
9617766-2011-00936
Event Type
Malfunction
Date Received
April 26, 2011
Date of Event
March 30, 2011
Report Date
April 26, 2011
Manufacturer
GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD (GE PRIVATE)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MANUFACTURER'S SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SINGLE BOARD COMPUTER WAS UPGRADED AND THE SYSTEM SOFTWARE WAS RELOADED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WERE NO IMAGES ON THE 8800 SYSTEM. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8800 FLUOROSCOPIC X-RAY JAA GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD (GE PRIVATE) 8800