FDA Adverse Event Death Summary report: N

CONSULTA CRT-D

MDR report key: 2090399 · Received May 16, 2011

Report

Report Number
2647346-2011-00690
Event Type
Death
Date Received
May 16, 2011
Date of Event
March 23, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S084
Removal / Correction Number
ASKU
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: (B)(4): THE DEVICE WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED), THE OUTER INSULATION WAS MELTED, THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR WAS DISTORTED. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THE DEFIBRILLATION CONDUCTOR HAD OVERSTRESS FRACTURE, THE OUTER INSULATION WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THERE WAS APPARENT EXPLANT DAMAGE. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

A CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND THREE LEADS WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. INFORMATION IDENTIFIED IN THE MANUFACTURE'S DATA BASE INDICATED THE PATIENT DIED APPROXIMATELY SEVEN MONTHS POST THE IMPLANT OF THE CRT-D SYSTEM. CAUSE OF DEATH WAS REQUESTED AND NOT RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONSULTA CRT-D IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. D224TRK ASKU

Patients

Seq Age Sex Outcome Treatment
1 70 YR Death