PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2011-03501
- Event Type
- Injury
- Date Received
- May 16, 2011
- Date of Event
- April 11, 2011
- Report Date
- April 24, 2011
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE SECOND PROGLIDE IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER. A CUFF MISS (NO SUTURE RETRIEVED/ATTACHED) CAN OCCUR DUE TO A NUMBER OF FACTORS INCLUDING, BUT NOT LIMITED TO, NEEDLE DEFLECTION DURING PLUNGER DEPLOYMENT DUE TO INTERACTION WITH HUMAN TISSUE, AGGRESSIVE AND FAST DEPLOYMENT OF THE PLUNGER OR A FAILURE TO MAINTAIN A STABLE POSITION OF THE DEVICE WITH RESPECT TO THE TISSUE TRACT DURING NEEDLE DEPLOYMENT. THIS MAY HAVE BEEN A CONTRIBUTING FACTOR TO THIS EVENT, BUT COULD NOT BE CONFIRMED. THE RETURN OF THE DEVICE MAY HAVE AIDED THE INVESTIGATION IN DETERMINING A CAUSE FOR THE EXPERIENCED EVENT. TO ENSURE THIS TYPE OF EXPERIENCE IS NOT A RESULT OF A POTENTIAL PRODUCT RELATED DEFICIENCY, IN PROCESS TESTING OF DEVICES IS PERFORMED TO ASSURE THERE IS NO NEEDLE DEFLECTION, WHICH MAY CAUSE THE EVENT REPORTED IN THIS CASE. A SAMPLING OF FINISHED DEVICES IS ALSO TESTED TO VERIFY THE FUNCTIONALITY OF THE DEVICE. A REVIEW OF THE FINISHED PRODUCT LOT HISTORY RECORD DID NOT REVEAL ANY MANUFACTURING RELATED NON-CONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT. BASED ON THE INFORMATION RECEIVED WITH THIS INCIDENT AND WITHOUT THE PRODUCT TO EXAMINE, A DEFINITIVE CAUSE FOR THE REPORTED EXPERIENCE COULD NOT BE DETERMINED.
IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED. THE DEVICE WAS REMOVED AND A SECOND PROGLIDE DEVICE WAS USED WITH THE SAME RESULTS. MANUAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 950166H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |