FDA Adverse Event Malfunction Summary report: N

CERENOVUS ENTERPRISE

MDR report key: 20903624 · Received December 11, 2024

Report

Report Number
3008114965-2024-01163
Event Type
Malfunction
Date Received
December 11, 2024
Date of Event
November 26, 2024
Report Date
December 27, 2024
Manufacturer
MEDOS INTERNATIONAL SARL
Product Code
NJE
UDI-DI
10886704043973
PMA / PMN Number
H60001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER¿S REF. NO: (B)(4). INFORMATION REGARDING PATIENT IDENTIFIER, DATE OF BIRTH, AGE, SEX, GENDER, WEIGHT, RACE, AND ETHNICITY WERE NOT PROVIDED. SECTION E.1: THE INITIAL REPORTER PHONE: (B)(6). SECTION H.3: THE DEVICE IS AVAILABLE TO BE RETURNED FOR EVALUATION AND TESTING. HOWEVER, IT HAS NOT BEEN RECEIVED TO DATE. IF THE DEVICE RETURNS, A DEVICE INVESTIGATION WILL BE PERFORMED. LAKE REGION MEDICAL PERFORMED A REVIEW OF THE DEVICE HISTORY RECORDS RELATIVE TO THE MANUFACTURING, INSPECTION, AND PACKAGING OF THE LOT 8962456. THE HISTORY RECORD INDICATES THIS PRODUCT WAS FINAL INSPECTION TESTED AT LAKE REGION MEDICAL AND WAS DETERMINED TO BE ACCEPTABLE. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY JOHNSON & JOHNSON MEDTECH, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, JOHNSON & JOHNSON MEDTECH, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. MISSING INFORMATION FROM THIS REPORT IS IDENTIFIED AS BLANK; THIS INFORMATION WAS NOT PROVIDED IN THE REPORTED EVENT OR AVAILABLE AT THE TIME OF REPORT SUBMISSION. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Additional Manufacturer Narrative · 0

MANUFACTURER¿S REF. NO: (B)(4). THE PURPOSE OF THIS MDR SUBMISSION IS TO REPORT THAT THE PRODUCT ANALYSIS LAB RECEIVED THE COMPLAINT DEVICE ON 23-DEC-2024. A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED ONCE THE PRODUCT INVESTIGATION HAS BEEN COMPLETED. UPDATED SECTIONS: B.4, D.9, G.3, G.6. H.2, H.3, H.6, AND H.11. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Additional Manufacturer Narrative · 0

MANUFACTURER¿S REF. NO: (B)(4). THE PURPOSE OF THIS MDR SUBMISSION IS TO REPORT THE INVESTIGATIONAL FINDING OF THE RETURNED DEVICE. THE COMPLAINT PRODUCT WAS RETURNED AND RECEIVED FOR EVALUATION AND ANALYSIS. THE INVESTIGATION IS DOCUMENTED BELOW. INVESTIGATION SUMMARY: A NON-STERILE 4.5MM X 14MM ENTERPRISE VASCULAR RECONSTRUCTION DEVICE WAS RECEIVED CONTAINED IN THE DECONTAMINATION POUCH. VISUAL INSPECTION WAS PERFORMED. AS DESCRIBED IN THE EVENT DESCRIPTION, THE STENT COMPONENT WAS DETACHED FROM THE DELIVERY SYSTEM. NO DAMAGES WERE NOTED. MICROSCOPIC INSPECTION WAS PERFORMED ON THE STENT COMPONENT. IT WAS OBSERVED TO BE IN GOOD CONDITION; THERE WAS NO STRUCTURAL DAMAGE (I.E., NO BROKEN STRUTS, NO KINKS); ALSO, IT WAS NOTED FULLY EXPANDED, BOTH ENDS CAN BE NOTED AS COMPLETELY FLARED. THE DIMENSIONS OF THE DELIVERY WIRE WERE MEASURED, AND ALL MEASUREMENTS WERE FOUND TO BE WITHIN SPECIFICATION, INCLUDING THOSE SPECIFICATIONS THAT CONTROL THE ATTACHMENT AND DELIVERY OF THE STENT. THEREFORE, DEVICE FAILURE IS NOT SUSPECTED TO BE A CONTRIBUTING FACTOR. THE ISSUE REPORTED REGARDING THE STENT BECOMING IMPEDED IN THE MICROCATHETER COULD NOT BE EVALUATED THROUGH FUNCTIONAL TESTING DUE TO THE DETACHED CONDITION OF THE STENT. IN ORDER TO PERFORM A FUNCTIONAL TEST, THE STENT MUST BE ATTACHED TO THE DELIVERY WIRE AND INSIDE THE INTRODUCER; HOWEVER, BASED ON THE DETACHED CONDITION, THE ISSUE REPORTED REGARDING THE PREMATURE SEPARATION OF THE STENT FROM THE DELIVERY WIRE IS CONFIRMED. NONE OF THE RETURNED COMPONENTS PRESENT DAMAGES THAT SUGGEST THAT THEY WERE FORCIBLY ADVANCED. WITH THE LIMITED INFORMATION AVAILABLE, A CONCLUSIVE CAUSE CANNOT BE DETERMINED; HOWEVER, IT IS POSSIBLE THAT CLINICAL AND PROCEDURAL FACTORS, INCLUDING DEVICE MANIPULATION AND OPERATOR'S TECHNIQUE, MAY HAVE CONTRIBUTED TO THE REPORTED FAILURE. AT THIS TIME, THERE IS NO EVIDENCE TO SUPPORT THAT THE ISSUE REPORTED IN THE COMPLAINT IS A RESULT OF A DEFECT INHERENTLY RELATED TO THE DEVICE. LAKE REGION MEDICAL PERFORMED A REVIEW OF THE DEVICE HISTORY RECORDS RELATIVE TO THE MANUFACTURING, INSPECTION, AND PACKAGING OF THE LOT 8962456. THE HISTORY RECORD INDICATES THIS PRODUCT WAS FINAL INSPECTION TESTED AT LAKE REGION MEDICAL AND WAS DETERMINED TO BE ACCEPTABLE. AS PART OF JOHNSON & JOHNSON MEDTECH QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. IT SHOULD BE NOTED THAT PRODUCT FAILURE COULD BE CAUSED BY MULTIPLE FACTORS. THE INSTRUCTIONS FOR USE (IFU) DOES CONTAIN THE FOLLOWING RECOMMENDATION: ¿ DO NOT APPLY UNDUE FORCE IF RESISTANCE IS ENCOUNTERED AT ANY POINT DURING STENT MANIPULATION. WITHDRAW THE UNIT AND ADVANCE TO A NEW ONE. BASED ON THE MANUFACTURING DOCUMENTATION REVIEW, THERE IS NO INDICATION THAT THE EVENT IS RELATED TO THE DEVICE MANUFACTURING PROCESS. AS PART OF THE POST MARKET SURVEILLANCE PROGRAM, INFORMATION FROM THIS COMPLAINT IS TRENDED FOR STATISTICAL SIGNALS AND CORRECTIVE / PREVENTIVE ACTION MAY BE TRIGGERED AT A LATER TIME. SINCE THERE WAS NO EVIDENCE TO SUGGEST THE EVENT WAS RELATED TO A MANUFACTURING OR DESIGN ISSUE, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME. UPDATED SECTIONS: B.4, G.3, G.6. H.2, H.3, H.6, AND H.11. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Description of Event or Problem · 0

THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A STENT-ASSISTED ENDOVASCULAR EMBOLIZATION OF AN ANTERIOR CEREBRAL COMMUNICATING ARTERY ANEURYSM, THE 4.5MM X 14MM ENTERPRISE VASCULAR RECONSTRUCTION DEVICE (ENC451412 / 8962456) WAS IMPEDED IN THE DISTAL END OF THE CONCOMITANT 150CM X 5CM PROWLER SELECT PLUS MICROCATHETER (606S255X / 31316405) AND COULD NOT PASS THROUGH THE MICROCATHETER. THE PHYSICIAN REMOVED THE STENT AND THE MICROCATHETER FROM THE PATIENT AND SWITCHED TO NEW DEVICES TO COMPLETE THE PROCEDURE. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. ON 10-DEC-2024, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE PATIENT IS A 77-YEAR-OLD MALE. THE PROCEDURE WAS CONFIRMED TO BE A STENT-ASSISTED EMBOLIZATION OF AN ANTERIOR CEREBRAL COMMUNICATING ARTERY ANEURYSM. AN ADEQUATE, CONTINUOUS FLUSH WAS MAINTAINED THROUGH THE MICROCATHETER. WHEN THE STENT WAS REMOVED FROM THE PATIENT, IT WAS NO LONGER STILL ON THE DELIVERY WIRE. THE STENT / STENT DELIVERY SYSTEM DID NOT APPEAR DAMAGED. THERE WERE NO DAMAGES NOTED ON THE MICROCATHETER. THE REPLACEMENT STENT WAS ANOTHER 4.5MM X 14MM ENTERPRISE VASCULAR RECONSTRUCTION DEVICE (ENC451412). THE REPLACEMENT MICROCATHETER WAS ANOTHER 150CM X 5CM PROWLER SELECT PLUS MICROCATHETER (606S255X). THE INFORMATION CONFIRMED THERE WAS NO NEGATIVE PATIENT IMPACT AND NO DELAY IN THE PROCEDURE AS A RESULT OF THE REPORTED ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2597754 CERENOVUS ENTERPRISE INTRACRANIAL NEUROVASCULAR STENT NJE MEDOS INTERNATIONAL SARL 8962456 10886704043973

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown PROWLER SELECT PLUS 150/5CM