FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2090353 · Received May 16, 2011

Report

Report Number
1423500-2011-06042
Event Type
Malfunction
Date Received
May 16, 2011
Date of Event
April 25, 2011
Report Date
April 25, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE IS NOT AVAILABLE FOR EVALUATION; THEREFORE, BAXTER CANNOT DETERMINE THE ROOT CAUSE. SINCE THE LOT NUMBER IS UNKNOWN A BATCH REVIEW WILL NOT BE PERFORMED.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR THE SYSTEM ERROR 2240 (AIR IN LINE) WAS NOT CONFIRMED. A ROOT CAUSE OF THE COMPLAINT WAS NOT DETERMINED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER AND REPORTED A SYSTEM ERROR 2240 (AIR IN LINE) ALARM THAT APPEARED ON THE HOME CHOICE (HC) MACHINE DURING DWELL 5 OF 5. THE HOME PATIENT (HP) HAS 4 BAGS CONNECTED AND THERE IS SOLUTION IN FINAL BAG ONLY. THERE WERE NO LEAKS AND NO PATIENT DISCONNECT. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) REFERRED THE HP TO THE NURSE. THE HP WILL COMPLETE WITH MANUAL SUPPLIES. THE ALARM CLEARED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO FURTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE