FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2090317 · Received May 16, 2011

Report

Report Number
1423500-2011-06029
Event Type
Malfunction
Date Received
May 16, 2011
Date of Event
April 22, 2011
Report Date
April 22, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS NOT RETURNED TO BAXTER, THEREFORE NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. A FOLLOW-UP REPORT WILL BE FILED SHOULD ANY SIGNIFICANT SUPPLEMENTAL INFORMATION BECOME AVAILABLE

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) WAS CONFIRMED; PER THE COMPLAINT INFORMATION THE CAUSE OF THE SE 2240 IS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE AFTER THE PATIENT DISCONNECTED THE PATIENT LINE FROM THE TRANSFER SET. A LABELING REVIEW OF THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE ISSUED WAS FOUND TO BE ADEQUATE FOR THE USE ERROR(S) IN THE COMPLAINT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR 2240 (AIR IN LINE) ALARM THAT APPEARED ON THE HOMECHOICE (HC) DISPLAY DURING DRAIN 4 OF 5. THE HOME PATIENT (HP) REPORTED HAVING DISCONNECTED AND THEN AFTER THE ALARM, HE RECONNECTED TO THE HC SYSTEM. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM AND THE POSSIBLE CAUSE, ASSISTED WITH TROUBLESHOOTING. PROPER PROCEDURES PER THE USER MANUAL WERE REVIEWED WITH THE HP. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.