FDA Adverse Event Malfunction Summary report: N

BD CTGCTV2

MDR report key: 20902537 · Received December 11, 2024

Report

Report Number
3007420875-2024-00180
Event Type
Malfunction
Date Received
December 11, 2024
Date of Event
October 23, 2024
Report Date
March 18, 2025
Manufacturer
GENEOHM SCIENCES CANADA, INC. (BD DIAGNOSTICS)
Product Code
LSL
UDI-DI
00382904439798
PMA / PMN Number
K210585
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D.2. ADDITIONAL MEDICAL DEVICE TYPES: MKZ, OUY, QEP . A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

AFTER FURTHER EVALUATION, THE PREVIOUS MFR REPORT # 3007420875-2024-00180 WAS SUBMITTED AGAINST WRONG PRODUCT. THIS REPORT IS TO BE REPLACE BY MFR REPORT # 1119779-2025-00272.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE OF BD CTGCTV2 ON BD MAX INSTRUMENT, 8 SAMPLES WERE FALSE POSITIVE FOR BOTH CHLAMYDIA TRACHOMATIS AND NEISSERIA GONORRHOEAE. AFTER INSPECTING THE REAGENTS AND INSTRUMENT, IT WAS VISIBLE THAT A "BLOODY SLIME STRING" FROM A SAMPLE WAS DRAGGED OVER THE PROCESS AREA AND CONTAMINATED THE SAMPLES. CLEANING AND ENVIRONMENTAL TESTING WERE PERFORMED. SAMPLES WERE RERUN AND WERE NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE OF BD CTGCTV2 ON BD MAX INSTRUMENT, 8 SAMPLES WERE FALSE POSITIVE FOR BOTH CHLAMYDIA TRACHOMATIS AND NEISSERIA GONORRHOEAE. AFTER INSPECTING THE REAGENTS AND INSTRUMENT, IT WAS VISIBLE THAT A "BLOODY SLIME STRING" FROM A SAMPLE WAS DRAGGED OVER THE PROCESS AREA AND CONTAMINATED THE SAMPLES. CLEANING AND ENVIRONMENTAL TESTING WERE PERFORMED. SAMPLES WERE RERUN AND WERE NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2573616 BD CTGCTV2 MULTIPLE-TYPE SEXUALLY TRANSMITTED PATHOGEN NUCLEIC ACID IVD, KIT, NUCLEIC ACID LSL GENEOHM SCIENCES CANADA, INC. (BD DIAGNOSTICS) 00382904439798

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown