BD CTGCTV2
Report
- Report Number
- 3007420875-2024-00180
- Event Type
- Malfunction
- Date Received
- December 11, 2024
- Date of Event
- October 23, 2024
- Report Date
- March 18, 2025
- Manufacturer
- GENEOHM SCIENCES CANADA, INC. (BD DIAGNOSTICS)
- Product Code
- LSL
- UDI-DI
- 00382904439798
- PMA / PMN Number
- K210585
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
D.2. ADDITIONAL MEDICAL DEVICE TYPES: MKZ, OUY, QEP . A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
AFTER FURTHER EVALUATION, THE PREVIOUS MFR REPORT # 3007420875-2024-00180 WAS SUBMITTED AGAINST WRONG PRODUCT. THIS REPORT IS TO BE REPLACE BY MFR REPORT # 1119779-2025-00272.
IT WAS REPORTED THAT DURING USE OF BD CTGCTV2 ON BD MAX INSTRUMENT, 8 SAMPLES WERE FALSE POSITIVE FOR BOTH CHLAMYDIA TRACHOMATIS AND NEISSERIA GONORRHOEAE. AFTER INSPECTING THE REAGENTS AND INSTRUMENT, IT WAS VISIBLE THAT A "BLOODY SLIME STRING" FROM A SAMPLE WAS DRAGGED OVER THE PROCESS AREA AND CONTAMINATED THE SAMPLES. CLEANING AND ENVIRONMENTAL TESTING WERE PERFORMED. SAMPLES WERE RERUN AND WERE NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.
IT WAS REPORTED THAT DURING USE OF BD CTGCTV2 ON BD MAX INSTRUMENT, 8 SAMPLES WERE FALSE POSITIVE FOR BOTH CHLAMYDIA TRACHOMATIS AND NEISSERIA GONORRHOEAE. AFTER INSPECTING THE REAGENTS AND INSTRUMENT, IT WAS VISIBLE THAT A "BLOODY SLIME STRING" FROM A SAMPLE WAS DRAGGED OVER THE PROCESS AREA AND CONTAMINATED THE SAMPLES. CLEANING AND ENVIRONMENTAL TESTING WERE PERFORMED. SAMPLES WERE RERUN AND WERE NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.