FDA Adverse Event Malfunction Summary report: N

UNIVERSAL HIGH TORQUE DRILL

MDR report key: 2090236 · Received April 19, 2011

Report

Report Number
1811755-2011-01308
Event Type
Malfunction
Date Received
April 19, 2011
Date of Event
March 23, 2011
Report Date
March 23, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DRILL WAS RECEIVED BY THE U.S. MFR AND THE COMPLAINT WAS CONFIRMED. INTERNAL COMPONENTS WERE EVALUATED AND CORROSION WAS DISCOVERED WITHIN THE ROTOR ASSEMBLY. THE PRESENCE OF CORROSION CAN LEAD TO INCREASED FRICTION BETWEEN ROTATING COMPONENTS, RESULTING IN HEAT BEING GENERATED. ADDITIONALLY, A CUT WAS DISCOVERED IN THE POWER CORD, WHICH IS A COMPONENT OF THE MOTOR ASSEMBLY. A DAMAGED MOTOR ASSEMBLY CAN ALSO RESULT IN HEAT BEING GENERATED, WHEN THE DRILL IS OPERATED. THE MOTOR ASSEMBLY AND ROTOR ASSEMBLY WERE REPLACED, AND ADD'L PREVENTATIVE MAINTENANCE WAS ALSO PERFORMED. THE DRILL WAS RETURNED TO THE USER FACILITY.

Description of Event or Problem · 1

IT WAS REPORTED BY A MFR TECHNICAL SPECIALIST, THAT THE DRILL HEATED UP. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNIVERSAL HIGH TORQUE DRILL ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK