FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 2090221 · Received May 16, 2011

Report

Report Number
1423500-2011-05989
Event Type
Malfunction
Date Received
May 16, 2011
Date of Event
March 31, 2011
Report Date
April 21, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS IS REPORT 3 OF 3 ASSOCIATED WITH THIS DEVICE. THE DEVICE HAS BEEN RECEIVED AT BAXTER FOR EVALUATION. HOWEVER, THE EVALUATION IS NOT COMPLETE AT THIS TIME. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE HOMECHOICE (HC) DEVICE WAS RECEIVED FOR EVALUATION AT THE PRODUCT ANALYSIS LAB (PAL). A REVIEW OF THE DEVICE LOGS CONFIRMED THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT. THE CAUSE OF THE IIPV FOUND IN THE LOG WAS DETERMINED TO BE INSUFFICIENT DRAIN DUE TO USE ERROR: THE TIDAL ULTRAFILTRATE (UF) REMOVAL WAS SET TOO LOW. A LABELING REVIEW OF THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE ISSUED WAS FOUND TO BE SUFFICIENT. A SERVICE HISTORY REVIEW (SHR) REVEALED THAT NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE IIPV. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

DURING A REVIEW OF THE LOGS OF A RETURNED HOMECHOICE (HC) MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS IDENTIFIED ON (B)(6) 2011 DURING DRAIN CYCLE 3. THE PATIENT'S LARGEST PRESCRIBED FILL VOLUME (LPFV) WAS 2000 ML AND THE DRAIN VOLUME WAS 3552 ML, WHICH MET IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. PRODUCT SURVEILLANCE CONTACTED THE NURSE AND NOTIFIED THE NURSE OF THE INCIDENTS OF IIPV THAT WERE FOUND DURING THE DEVICE LOG REVIEW.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 33 YR