FDA Adverse Event Injury Summary report: N

PRIMARY TRITANIUM HEMI CLUSTER HOLE CUP 54MM

MDR report key: 2090124 · Received May 5, 2011

Report

Report Number
2249697-2011-00656
Event Type
Injury
Date Received
May 5, 2011
Date of Event
September 3, 2010
Report Date
April 9, 2011
Manufacturer
STRYKER ORTHOPAEDICS MAHWAH
Product Code
LPH
PMA / PMN Number
K081171
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION. CATALOG NUMBERS AND LOT CODES OF OTHER DEVICES LISTED IN THIS REPORT: CAT # 690-00-22E, LOT # MJDJTD, DESCRIPTION: TRIDENT 0 DEG CONSTRAINED INSERT 22E. THIS IS THE SAME PT/EVENT AS MFR # 9616680-2011-00280.

Description of Event or Problem · 1

IT WAS REPORTED: PT ORIGINALLY WENT IN FOR A THA ON (B)(6) 2010. A 54M TRITANIUM SHELL USED, WITH SCREWS AND A 32 'E 0 DEGREE X3 LINER. EXETER 44 #2 STEM WAS ALSO IMPLANTED (ALONG WITH 32MM LFIT V40 HEAD AND 12MM BONE PLUG). THREE TO FOUR WEEKS POST-OP, PT DISLOCATED HIP (PT WAS POTENTIALLY PUSHING RECOVERY TOO HARD, BY REACHING DOWN TO TOUCH TOES AND IN THE PROCESS TORE HER GLUTEAL MUSCLES). THIS WAS FOLLOWED BY 1-2 MORE DISLOCATIONS, BEFORE A REVISION ON (B)(6) 2010 TO PUT IN A CONSTRAINED LINER. SHELL REMAINED IN SITU FROM THE REVISION, WITH ORIGINAL LINER REMOVED AND CONSTRAINED LINER INSERTED. GLUTEAL MUSCLES WERE ALSO RE-ATTACHED USING MITEK ANCHORS. SIX WEEK POST-REVISION X-RAY WAS ALL GOOD, HOWEVER PT FELT CLICKING IN HIP AND GRABBING PAIN IN (B)(6) 2010. RECEIVED AN X-RAY ON (B)(6) 2011, WHERE ANOTHER REVISION WAS RECOMMENDED. RE-REVISION WAS PERFORMED ON (B)(6) 2011, WHERE THE TRIATANIUM SHELL FROM THE ORIGINAL OPERATION ((B)(6) 2010) AND CONSTRAINED LINER (FROM (B)(6) 2010 REVISION) WAS REMOVED AND REPLACED WITH A TRIATANIUM 56MM 'E' SHELL (ALONG WITH 3X SCREWS) AND TRIDENT 0 DEGREE CONSTRAINED LINER 'E' (AND ASSOCIATED 22MM HEAD). THE ORIGINAL EXETER STEM REMAINED IN SITU".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PRIMARY TRITANIUM HEMI CLUSTER HOLE CUP 54MM IMPLANT LPH STRYKER ORTHOPAEDICS MAHWAH NA MHLEWV

Patients

Seq Age Sex Outcome Treatment
1 78 YR Other| R