COBAS 8000 C 701 MODULE
Report
- Report Number
- 1823260-2024-03604
- Event Type
- Malfunction
- Date Received
- December 10, 2024
- Date of Event
- November 26, 2024
- Report Date
- December 10, 2024
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K100853
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE GLUCOSE REAGENT LOT NUMBER WAS 812051. THE CHOLESTEROL REAGENT LOT NUMBER AND THE REAGENT EXPIRATION DATES WERE REQUESTED, BUT NOT PROVIDED. THE FIELD SERVICE ENGINEER FOUND DAMAGED CELL RINSE ASSEMBLY TUBING. THE TUBING WAS REPLACED. A VALVE AND NOZZLE WERE ALSO REPLACED. MAINTENANCE WAS PERFORMED ON THE ANALYZER. THE INVESTIGATION DETERMINED THE SERVICE ACTIONS RESOLVED THE ISSUE AND THE ISSUE WAS RELATED TO SERVICE MAINTENANCE.
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR AN UNSPECIFIED NUMBER OF PATIENT SAMPLES TESTED WITH GLUCOSE HK GEN. 3 ON A COBAS 8000 C 701 MODULE. THE CUSTOMER PROVIDED AN EXAMPLE OF ONE PATIENT SAMPLE WITH DISCREPANT GLUCOSE RESULTS. A SECOND PATIENT SAMPLE ALSO HAD DISCREPANT CHOLESTEROL RESULTS WHEN TESTED ON THE C 701 ANALYZER. THE SAMPLE INITIALLY RESULTED IN A GLUCOSE VALUE OF 279.9 MG/L. THE PATIENT WAS CALLED TO THE HOSPITAL WHERE A RAPID TEST FOR GLUCOSE DETERMINATION WAS PERFORMED. THE RAPID TEST GLUCOSE VALUE WAS WITHIN THE NORMAL REFERENCE RANGE. THE PATIENT COMPLAINED ABOUT THE INITIAL SAMPLE VALUE BASED ON THIS AND THE SAMPLE WAS REPEATED. THE REPEAT RESULT ON (B)(6) 2025 WAS 88.2 MG/DL. THE REPEAT VALUE AGREED WITH THE RAPID TEST RESULT. THE SECOND SAMPLE INITIALLY RESULTED IN A CHOLESTEROL VALUE OF 84 MG/DL AND IT REPEATED AS 186 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1677868 | COBAS 8000 C 701 MODULE | CLINICAL CHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR | Female |