FDA Adverse Event Malfunction Summary report: N

PATIENT INFORMATION CENTER IX

MDR report key: 20894689 · Received December 10, 2024

Report

Report Number
9610816-2024-000954
Event Type
Malfunction
Date Received
December 10, 2024
Date of Event
November 21, 2024
Report Date
January 30, 2025
Manufacturer
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
Product Code
MHX
UDI-DI
00884838121782
PMA / PMN Number
K211900
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A PHILIPS REMOTE SERVICE ENGINEER (RSE) INTERVIEWED THE CUSTOMER, AND IT WAS DISCOVERED THERE WAS AN ISSUE WITHIN THE BEDSIDE AREA DESPITE A REBOOT OF THE PLANT. THE RSE CONFIRMED THE ISSUE TO BE A BED-TO-BED OVERVIEW FAILURE. AFTER A REBOOT OF THE CENTRAL STATION, THE BED-TO-BED OVERVIEW WAS FUNCTIONAL AGAIN. NO FURTHER INVESTIGATION OR ACTION IS WARRANTED AT THIS TIME. UPDATED INFORMATION FINDS THIS RECORD TO BE NOT REPORTABLE. ALARMS ARE OCCURRING BOTH ON THE OTHER PATIENTS MONITOR AND ON THE PIC IX. BED TO BED OVERVIEW AT THE INTELLIVUE PATIENT MONITOR IS A SECONDARY ALARM NOTIFICATION SYSTEM AND IS NOT INTENDED FOR PRIMARY NOTIFICATION OF ALARMS, PHYSIOLOGICAL DATA, OR DEMOGRAPHIC DATA. THE INSTRUCTIONS FOR USE (IFU) FOR PIC IX C.03 CONTAINS UPDATED LABELING REGARDING CAREGROUPS AND INTENDED USE.

Additional Manufacturer Narrative · 0

PHILIPS IS IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION CONCERNING THIS EVENT AND THE COMPLAINT IS STILL UNDER INVESTIGATION. A FINAL REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION IS COMPLETE. E1: REPORTER INSTITUTION PHONE NUMBER: (B)(6).

Description of Event or Problem · 0

IT WAS REPORTED THERE WAS A PROBLEM WITH INTERMITTENT ALARMS THEY WERE ABSENT AND SOLVED BY A REBOOT, BUT THE PROBLEM RETURNED SHORTLY AFTERWARDS. IT IS UNKNOWN IF THE DEVICE WAS IN USE AT TIME OF EVENT, AND THERE WAS NO ADVERSE EVENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2536727 PATIENT INFORMATION CENTER IX PATIENT INFORMATION CENTER IX MHX PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH 866389 00884838121782

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown