FDA Adverse Event Malfunction Summary report: N

AUTOSOFT XC

MDR report key: 20876427 · Received December 9, 2024

Report

Report Number
3003442380-2024-33953
Event Type
Malfunction
Date Received
December 9, 2024
Date of Event
November 3, 2024
Report Date
November 4, 2025
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244016620
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 01 - (B)(4) - MDR 3003442380-2024-33953. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. REVISION 21 OF RECORD (B)(4) DOES NOT REQUIRE A TYPE 2 REPORTABLE COMPLAINT TO INITIATE A CHILD INVESTIGATION. THIS CHILD INVESTIGATION WAS CREATED AND SUBSEQUENTLY CLOSED TO DOCUMENT THE DEVICE HISTORY RECORD (DHR) REVIEW, WHICH WAS NECESSARY TO ADVANCE THE PARENT RECORD TO THE REVIEW AND SUMMARY. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6003996 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DHR REVIEW: THE LOT 6003996 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 108 AND MANUFACTURED IN THE LINE 7 ON (B)(6) 2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING OF TUBING OF THE LOT 3K03852 WAS MANUFACTURED ACCORDING TO THE WI VERSION 57 AND MANUFACTURED IN THE MACHINE SC01, ON (B)(6) 2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING OF TUBING OF THE LOT 3K03853 WAS MANUFACTURED ACCORDING TO THE WI VERSION 57 AND MANUFACTURED IN THE MACHINE SC03, ON (B)(6) 2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT MALFUNCTION CODE, THEREFORE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Additional Manufacturer Narrative · 0

INITIAL AND FINAL (B)(4) - DEVICE 2 OF 2.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED TWO INFUSION SETS TUBING DETACHED EVENTS ON (B)(6) 2024 FROM CONNECTOR. INFUSION SETS WERE USED FOR 2.5 HOUR TO 1 DAY FOR BOTH EVENTS. PATIENT REPLACED INFUSION SETS AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
435127 AUTOSOFT XC UNO INSET I 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL A/S 1001680 6003996 05705244016620

Patients

Seq Age Sex Outcome Treatment
1 57 YR Male