FDA Adverse Event Malfunction Summary report: N

R SERIES

MDR report key: 20863515 · Received December 6, 2024

Report

Report Number
1220908-2024-04324
Event Type
Malfunction
Date Received
December 6, 2024
Report Date
November 15, 2024
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
UDI-DI
00847946017446
PMA / PMN Number
P160022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS EVALUATED BYZ OLL MEDICAL CORPORATION. THE CUSTOMER'S REPORT WAS DUPLICATED AND ATTRIBUTED TO AN INTEGRATED CIRCUIT ON THE DIGITAL BOARD. THE DIGITAL BOARD WAS REPLACED TO RESOLVE THE REPORT. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Additional Manufacturer Narrative · 0

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A SUPPLEMENTAL REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Description of Event or Problem · 0

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2121419 R SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION 30001004101110013 NA 00847946017446

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown