FDA Adverse Event
Injury
Summary report: N
OCTRODE LEAD KIT, 60CM LENGTH
MDR report key: 20861373
·
Received December 6, 2024
Report
- Report Number
- 1627487-2024-12466
- Event Type
- Injury
- Date Received
- December 6, 2024
- Date of Event
- November 19, 2024
- Report Date
- December 6, 2024
- Manufacturer
- ABBOTT MEDICAL
- Product Code
- LGW
- UDI-DI
- 05415067017246
- PMA / PMN Number
- P010032
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
DATE OF EVENT IS ESTIMATED. ADDITIONAL COMPONENTS POTENTIALLY INVOLVED IN THE EVENT INCLUDE: COMMON DEVICE NAME: SCS LEAD, MODEL: 3186, UDI: (B)(4), SERIAL: (B)(6), BATCH: 8031786. THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. BASED ON THE INFORMATION RECEIVED, A SINGLE DEFINITIVE ROOT CAUSE FOR THE ISSUE ENCOUNTERED WAS UNABLE TO BE CONCLUSIVELY DETERMINED.
Description of Event or Problem · 0
IT WAS REPORTED ONE OF PATIENT'S LEAD HAD HIGH IMPEDANCES ON ALL CONTACTS. INVESTIGATION WAS UNABLE TO DETERMINE WHICH LEAD ATTRIBUTED TO THE ISSUE. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD WAS EXPLANTED AND REPLACED. THERAPY WAS RESTORED POST-OP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2245467 | OCTRODE LEAD KIT, 60CM LENGTH | SCS LEAD | LGW | ABBOTT MEDICAL | 3186 | 8031786 | 05415067017246 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Female | Other | SCS ANCHOR 2X.| SCS IPG.| SCS LEAD. |