FDA Adverse Event Malfunction Summary report: N

CARESCAPE

MDR report key: 20859525 · Received December 5, 2024

Report

Report Number
2112667-2024-06508
Event Type
Malfunction
Date Received
December 5, 2024
Date of Event
November 7, 2024
Report Date
December 5, 2024
Manufacturer
DATEX-OHMEDA, INC.
Product Code
CBK
UDI-DI
00840682102346
PMA / PMN Number
K210384
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EG
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE ELECTRONIC FLOW CONTROL (EFC) VALVE WAS REPLACED TO RESOLVE THE ISSUE. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: HCS MADISON - 3030 OHMEDA DR, USA MADISON, WI 53718.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT OXYGEN OR FRESH GAS FLOW. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1377384 CARESCAPE VENTILATOR, CONTINUOUS, FACILITY USE CBK DATEX-OHMEDA, INC. R860 NA 00840682102346

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown