VIVASIGHT 2 DLT KIT 39 FR - LEFT
Report
- Report Number
- 1220828-2024-00025
- Event Type
- Injury
- Date Received
- December 5, 2024
- Date of Event
- November 6, 2024
- Report Date
- December 5, 2024
- Manufacturer
- AMBU A/S
- Product Code
- CBI
- UDI-DI
- 05707480145768
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- HI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
MEASUREMENT FROM END TO TIP OF ENDOTRACHEAL BEVEL WAS PERFORMED, IN ORDER TO EVALUATE IF THE BENDING IS WITHIN SPECIFICATIONS (15°-53°). THE ANGLE OF THE RETENTION SAMPLE WAS 60°, THUS OUTSIDE OF SPECIFICATIONS. CORRECTIVE ACTION HAD BEEN INITIATED TO INVESTIGATE THIS ISSUE AND THE PRELIMINARY INVESTIGATION SHOWED THAT THE BENDING WAS CAUSED BY MISHANDLING BY MANUFACTURING PERSONNEL, IN WHICH ADDITIONAL BENDING OF ENDOBRONCHIAL TIP OCCURRED AFTER STYLET INSERTION PROCESS. THERE IS AN ONGOING RECALL 'AMBU2024FA00001' WHICH WAS INITIATED DUE TO HYPERANGULATION OF THE DISTAL TIP. THE LOT NUMBER OF THE AFFECTED PRODUCT INVOLVED IN THIS INCIDENT CORRESPONDS TO THE LOT NUMBERS AFFECTED BY THE RECALL.
THE INCIDENT LEADING TO TONSIL INJURIES OCCURRED IN TWO SEPARATE PROCEDURES (THE SECOND INCIDENT IS COVERED BY 1220828-2024- 00024). A CRNA WAS TRYING TO INTUBATE WITH THE VIVASIGHT 2 DLT PRODUCT AND THEY HAD TO ADJUST THE DISTANCE OF TIP OF THE TUBE PAST THE VOCAL CORDS. THE CURVE AT THE END OF THE BRONCHIAL LUMEN WAS MORE ANGULATED THAN A TYPICAL ROBERT-SHAW DOUBLE LUMEN ENDOTRACHEAL TUBE, AND GOT MORE EASILY CAUGHT ON THE TRACHEAL RINGS. ONE COULD EITHER ADVANCE THE BRONCHIAL CUFF FURTHER PAST THE VOCAL CORDS WITH THE STYLET, OR IF THE STYLET WAS REMOVED AFTER THE TIP WAS IN THE TRACHEA, ONE COULD TURN THE TUBE 120-180 DEGREES LEFTWARD WHILE ADVANCING AND THEN RETURN IT TO 90 DEGREES ONCE PAST THE RING THAT THE TIP GOT HUNG UP ON.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2517686 | VIVASIGHT 2 DLT KIT 39 FR - LEFT | VIVASIGHT 2 DLT KIT 39 FR - LEFT | CBI | AMBU A/S | 412391000 | 1000989835 | 05707480145768 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |