FDA Adverse Event Malfunction Summary report: N

SYNCHRON® CX5 DELTA CLINICAL SYSTEM

MDR report key: 2083763 · Received May 10, 2011

Report

Report Number
2050012-2011-01604
Event Type
Malfunction
Date Received
May 10, 2011
Date of Event
April 14, 2011
Report Date
April 14, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K881495
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BCI FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED FOR THIS EVENT. FSE REPLACED THE WASH STATION TOWER AND VERIFIED REPAIR.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT THAT THE Y16 PORT FITTING OF THE SYNCHRON CX5 DELTA CLINICAL SYSTEM IS LOOSE AND LEAKS EACH TIME THE WASH TOWER PRIMES. NO REPORTS OF DEATH OR INJURY HAVE BEEN REPORTED IN CONNECTION WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHRON® CX5 DELTA CLINICAL SYSTEM CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. CX5 DELTA N/A

Patients

Seq Age Sex Outcome Treatment
1