FDA Adverse Event Injury Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2083762 · Received May 10, 2011

Report

Report Number
2531779-2011-03317
Event Type
Injury
Date Received
May 10, 2011
Date of Event
April 17, 2011
Report Date
April 17, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILLED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. THE REPORT IS UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.

Additional Manufacturer Narrative · 1

ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD ON (B)(6) 2011 FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/17/2011 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP HISTORY INDICATED THAT INSULIN DELIVERY TOTALS CORRECTLY REFLECTED PROGRAMMED VALUES. THERE WAS NO ACTIVITY OUTSIDE NORMAL USE OBSERVED IN THE PUMP HISTORY. THE PUMP WAS EXERCISED FOR 29 HOURS WITH NO INSULIN DELIVERY ISSUES. THERE WERE NO DEFECTS FOUND ON INVESTIGATION.

Description of Event or Problem · 1

THE REPORTER CLAIMED THAT HER BLOOD GLUCOSE WAS LOWERED TO "44 MG/DL" WITH SYMPTOMS OF SHAKY, SWEATY, AND WANTED TO SLEEP AFTER SHE TOOK BOLUS INSULIN FOR TWO BREAKFAST PIZZA AND ONLY CONSUMED ONE PIZZA. HER TARGET BLOOD GLUCOSE IS 120 MG/DL. THE PATIENT STAYED BELOW 120 MG/DL ON THE DAY OF CONCERN UNTIL SHE ATE TWO POP TARTS, AND SODA. DURING TROUBLESHOOTING, THE ANIMAS REPRESENTATIVE CONCLUDED THAT THE INSULIN PUMP IS WORKING ACCORDINGLY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT USED THE ANIMAS INSULIN PUMP, REPORTEDLY TOOK TOO MUCH INSULIN FOR THE MEAL SHE ATE AND SUBSEQUENTLY SUFFERED A SERIOUS INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 52 YR Life Threatening| R