FDA Adverse Event Injury Summary report: N

TENDRIL ST

MDR report key: 2083758 · Received May 10, 2011

Report

Report Number
2017865-2011-02913
Event Type
Injury
Date Received
May 10, 2011
Date of Event
February 2, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVN
PMA / PMN Number
P960013
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. FINAL ANALYSIS OF THE LEAD FOUND THAT THE PROXIMAL INSULATION WAS MELTED/DAMAGED AT SEVERAL PLACES AT A DISTANCE OF 20 CM FROM THE CONNECTOR PIN. THE DAMAGE FOUND IS CONSISTENT WITH THAT OCCURRING AT THE TIME OF EXPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING CHANGE OUT OF THE PATIENT'S PULSE GENERATOR, LEAD ABRASIONS WERE OBSERVED. THE LEAD WAS REMOVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL ST PERMANENT PACEMAKER ELECTRODE NVN ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1888TC/58 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention 1258T/75 (B)(4), 1999/52 (B)(4)