FDA Adverse Event Injury Summary report: N

TENDRIL ST

MDR report key: 2083738 · Received May 10, 2011

Report

Report Number
2017865-2011-02890
Event Type
Injury
Date Received
May 10, 2011
Date of Event
February 11, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVN
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED FINAL ANALYSIS FOUND THAT THE PROXIMAL INSULATION WAS ABRADED AT 18.5 CM FROM THE CONNECTOR PIN. ABRASIONS ARE INDICATIVE OF EXPOSURE TO CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD INSULATION EXHIBITED AN ANOMALY. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL ST PERMANENT PACEMAKER ELECTRODE NVN ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1888TC/52 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention 1888TC/58, (B)(4)